AA-7090 Atomic Absorption Spectrophotometer
Product Description
Model: AA-7090
Manufacturer: East-West Analysis EWAI
All-in-one unit for parallel installation of flame graphite furnaces with both deuterium lamps and Seemann background corrections
Magnetic field strength variable from 0.6T to 1.1T with 0.1T amplitude
With coded light function
Eight-lamp automatic turret, rapid automatic selection of elemental lamps, can truly realize the same project multi-element sequential determination
Wavelength setting: 180nm~900nm, computer controlled automatic wavelength fast scanning.
Original imported transverse heating graphite furnaceThe dual background correction modes of longitudinal AC Seeman and deuterium lamp are used to meet the analytical needs of different users;
Burning head: 10cm seam length, all titanium material, high salt and corrosion resistant;
The AC Seeman background correction magnetic field is continuously adjustable from 0.6T to 1.1T, reducing the influence of inversion effect on the analytical sensitivity;
The absorbance of 1 ppm copper was better than 0.2 Abs;
Temperature rise rate ≥2500℃/sec, adjustable;
Graphite Furnace Visualization SystemThe graphite furnace is used to optimize the temperature of the graphite furnace, to locate the injection needle, to observe the sample in the graphite tube and to optimize the graphite furnace temperature program;
Maximum working temperature of graphite furnace: 2700℃;
Graphite furnace gas-saving mode, improve the effective utilization of protective gas, extend the gas use time, reduce the cost of use;
Dual protection gas circuit for graphite furnace, prolonging the graphite tube life by controlling the inner gas circuit and auxiliary gas circuit;
Elemental lamp selection, wavelength selection, slit selection, combustion head height and flow rate setting, automatic ignition, atomizer switching, deuterium light path switching, graphite furnace power switch and other functions can be realized through software;
Deuterium lamp fine-tuning mechanism to adjust the deuterium lamp line to coincide with the elemental lamp line;
With laughing gas safety lock and related safety mechanismIt is safer to use with a laughing gas-acetylene flame.
Product Specification
Instrument type: Flame-graphite furnace
Monochrome element: Planar grating
Optical system: single beam
Detectors: photomultiplier tubes
Buckling background technology: Deuterium lamp + Seman buckling background
Resolution: 0.1nm
Sensitivity: Flame method: Cu: ≤0.02μg/mL; Graphite furnace method: Cd: ≤ 5.0×10-13g
Repeatability (RSD): Flame analysis: Cu ≤0.45%; Graphite furnace analysis: Cd ≤3%
Detection limit: Flame analysis: Cu≤0.003μg/mL; Graphite furnace analysis: Cd≤1.0×10-12g
Baseline stability: ≤0.002A/30min (static); ≤0.003A/30min (dynamic)
Product Features
Model AA-7090 Atomic Absorption Spectrophotometer is a flame and graphite furnace installed in parallel with the integrated equipment, with deuterium lamps and Seemann two kinds of background correction, equipped with the original imported transverse heating graphite furnace, and graphite furnace visualization system and gas saving mode. The instrument adopts eight-lamp automatic turret, which can quickly switch the elemental lamps, and at the same time realizes a variety of automation functions through software control. In addition, the instrument has high sensitivity, low detection limit and good baseline stability, and is equipped with safety mechanisms such as laughing gas safety lock.